65
Display Error code
Description
Unit mass error
The sample unit mass for the counting mode is too
light. Storing and using it for counting will cause a
counting error.
Add samples to reach the specified number and press
the PRINT key.
Pressing the PRINT key without adding samples will
shift the balance to the counting mode. But, to acquire
accurate weighing, be sure to add samples.
Balance internal error
If this error appears persistently, contact the local A&D
dealer.
Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and
calendar function will continue to work normally as long
as the AC adapter is connected to the balance.
To replace the backup battery, contact the local A&D
dealer.
EC, E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01
Undefined command error
An undefined command was received.
Confirm the command.
EC, E02
Not ready
A received command can not be processed.
e.g. The balance received a Q command, but not in the
weighing mode.
e.g. The balance received a Q command while
processing a RE-ZERO command.
Adjust the delay time to transmit a command.
EC, E03
Timeout error
If the timeout parameter is set to “
t-Up1
”, the balance
did not receive the next character of a command within
the time limit of one second.
Confirm the communication.
EC, E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC, E06
Format error
A command includes incorrect data.
e.g. The data is numerically incorrect.
Confirm the command.
EC, E07
Parameter setting error
The received data exceeds the range that the balance
can accept.
Confirm the parameter range of the command.
Other errors
If the errors described above can not be released or
other errors are displayed, contact the local A&D dealer.
Summary of Contents for FZ-1200CT
Page 35: ...33 9 7 Data Format Examples LF...
Page 36: ...34...
Page 74: ...72 20 EXTERNAL DIMENSIONS FZ 1200CT 700CT 500CT FX 1200CT 700CT 500CT Unit mm...
Page 80: ...78 MEMO...
Page 81: ...79 MEMO...
Page 82: ...80 MEMO...
Page 83: ......