19
5. SMART
RANGE
FUNCTION
The AD-4212D-32 is equipped with two ranges. The precision range has a higher resolution. The
standard range has a normal resolution.
The range switches automatically depending on the value displayed. Touching the RE-ZERO button
or sending the R command allows weighing in the precision range regardless of the tare value. The
range can be fixed to the standard range by changing the minimum weighing value using the
RE-ZERO button or R command.
1 Touch the RE-ZERO button or send the R
command.
The balance will start weighing, using the
precision range.
2 Place a container on the weighing pan.
The weight value exceeds the precision range
value and the balance will switch to the
standard range.
3 Touch the RE-ZERO button or send the R
command.
The balance will switch to the precision range.
4 Place a sample in the container.
The weight value is within the precision range
value and the balance will perform a weighing,
using the precision range.
Precision range/standard range
Model
Range used
Weighing range
Available minimum
weighing value
Precision range
*
0
g
to 5.1 g
0.000001g (0.001mg)
AD-4212D-32
Standard range
5.1 g to 32 g
0.00001g (0.01mg)
The factory setting is the precision range.
*
When the weight value exceeds 5.1 g while the minimum weighing value of 0.000001 g
(precision range) is used, the balance will switch to the minimum weighing value of
0.00001 g (standard range). By touching the RE-ZERO button or sending the R command,
the balance can weigh from that point up to 5.1 g, using the minimum weighing value of
0.000001 g (precision range).
Standard range
Precision range
Precision range
Precision range
Container
Weighing
pan
Sample