32
8.2. Commands
Command List
Note A command has a terminator added, that is specified using "
5if
,
Crlf
" of the function
table, and is sent to the balance.
Commands to query weighing data
C
Cancels the S or SIR command.
Q
Requests the weighing data immediately.
S
Requests the weighing data when stabilized.
SI
Requests the weighing data immediately.
SIR
Requests the weighing data continuously.
Commands to control the balance
CAL
Performs calibration using the internal weight.
EXC
Performs calibration using the external weight.
OFF
Goes into standby mode. (Weighing value output is not performed)
ON
Goes into weighing mode from standby mode.
P
Switches between weighing mode and standby mode.
PRT
Confirms operations at the calibration etc.
R
Performs re-zero.
#1
RNG
Switches the minimum display.
TR
Tares the balance.
#1
U
Changes the weighing speed.(Response speed)
Commands to request data
?ID
Outputs the identification number.
?PT
Outputs the tare value.
#1
?SN
Outputs the serial number of the balance.
?TN
Outputs the model name of the balance.
?TM
Outputs time.
?DT
Outputs date.
Command to set data
PT:
.
g
Sets the tare value.
#1
TM:
:
:
Sets time.
Example)
12:34:56
TM: 12/34/56
DT:
/
/
Sets date.
Example) October 1, 2015
DT: 15/10/01
ID:
/
/
Sets the ID number.
#1
Command
R
assumes the point as zero and sets the display to zero.
Command
TR
cancels the tare weight when the displayed value is greater than zero.
The canceled tare value can be read using the command
?PT
.