7.2. Details Of The Function Table
Class
Item and Parameter
Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by
response
adjustment.
0
Stable range is ±1 digit
1
±2
digit
#1
5t-b
Stability band width
2
Stable range is ±3 digits
The stabilization indicator
illuminates when the display
fluctuation is within the range.
off
OFF
ploc
Display lock function
on
ON
Display hold function to
measure the volume dispensed
from the pipette.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by
tracking zero drift.
0
5 times/second
#3
5pd
Display refresh rate
1
10 times/second
#3
Period to refresh the display
0
Point (.)
ba5fnc
Environment
Display
pnt
Decimal point
1
Comma (,)
Decimal point format
0
Normal mode
Outputs data by commands.
1
Auto print mode A
Reference = zero
2
Auto print mode B
Reference = last stable value
Outputs data when the
display is stable and
conditions of
ap-p
,
ap-b
and
the reference value are met.
prt
Data output mode
3
Stream mode /
Interval mode
0
Plus only
Displayed value > Reference
1
Minus only
Displayed value < Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed
value
0
10 digits
1
100 digits
dout
Data output
#4
ap-b
Auto print difference
2
1000 digits
Difference between reference
value and displayed value
#2
: Factory settings.
#1 The unit of minimum display is digit.
Example: If 1 mg display is selected touching the RANGE button for the AD4212D-301, 1 mg
is one digit.
#2 Usable minimum display of the balance is one digit.
Example: In gram display, one digit is 0.00001 mg for the AD4212D-302 and 0.0001 mg for the
AD4212D-301.
#3 Factory settings of the AD4212D-33 and AD4212D-32 is set to 5 times per second.
Factory settings of the AD4212D-302 and AD4212D-301 is set to 10 times per second.
#4 These items apply only for RS-232C output for the PC or PLC.