Caution
The balance may not transmit the data completely at the specified refresh rate,
depending on the baud rate or data added to the weighing data such as time, date and ID
number.
#4
These items apply only for RS-232C output for the PC or PLC.
7.3. Description Of The Class "Environment, Display"
Condition (
Cond
)
Cond
0
This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target mass, weighing a very light sample or when quick response weighing is
required.
Cond
2
This parameter is for stable weighing with slow response. Used to prevent a mass
value from drifting due to vibration or drafts.
Notes
In automatic response adjustment, this parameter is selected automatically.
Stability band width
(
5t-b
)
This item controls the width to regard a mass value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs.
The parameter influences the "Auto print mode". Currently displayed minimum display means 1
digit.
Example) If the RANGE button was touched to select the 1 mg display at the AD4212D-301, 1
mg means 1 digit.
5t-b
0
This parameter is used for sensitive response of the stabilization indicator. Used for
exact weighing.
5t-b
2
This parameter ignores slight fluctuations of a mass value. Used to prevent a mass
value from drifting due to vibration or drafts.
Zero tracking (
trc
)
This function tracks zero point drift caused by changes in the environment and stabilizes the zero
point. When the weighing data is only a few digits, turn the function off for accurate weighing.
trc
0
The tracking function is not used. Used for weighing a very light sample.
trc
1
The normal tracking function is used.
trc
2
The strong tracking function is used.
trc
3
The very strong tracking function is used. Used for stable zero display.
Display refresh rate (
5pd
)
The periodic time to refresh the display. This parameter influences "Baud rate", "Data output
pause" and the data output rate of "Stream mode".
Decimal point (
pnt
)
The decimal point format can be selected.