9. MAINTENANCE
9.1. Treatment Of The Balance
When cleaning, remove the weighing pan and pan support from the balance and attach the
waterproof cover for cleaning the balance.
Do not use organic solvents to clean the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or
repair.
Use the original packing material for transportation. When packing, fasten the waterproof cover
to the balance using the tape by following the instructions at the figure below.
Consider
"
2. UNPACKING THE BALANCE
" and “
3.PRECAUTIONS
” when operation the
balance.
9.2. Error Codes
Error code
Description
EC,E00
Internal error
If displaying this error continuously, contact your local A&D dealer to repair the
balance.
EC,E11
Stability error
The balance can not stabilize due to an environmental problem. Check around the
pan. Prevent vibration, drafts, temperature changes, static electricity and magnetic
fields, from influencing the balance.
EC,E16
Internal weight error
Applying the internal weight does not yield a change in the weight value as
specified. Confirm that there is nothing on the pan and perform the weighing
operation from the beginning again.
EC,E17
Internal weight error
The internal weight application mechanism does not function properly. Perform the
weighing operation from the beginning again.
EC,E20
Calibration weight error
The calibration weight is too heavy. Confirm the calibration weight value. Press the
1
CAL
1
key to return to the weighing mode.
2
1
3
Remove the weighing pan
and pan support
.
Waterproof cover for cleaning
Fasten the waterproof
cover to the balance using
tape.