41
10. SPECIFICATIONS
AD-4212D-33 AD-4212D-32 AD-4212D-302 AD-4212D-301
32 g
Weighing capacity
32 g
5.1 g
320 g
320 g
32.00008 g
Maximum display
32.000084 g
5.100009 g
320.00084 g
320.0084 g
0.01 mg
Minimum display
0.001 mg
0.001 mg
0.01 mg
0.1 mg
0.02 mg
Repeatability
(Standard
deviation)
0.002 mg
0.004 mg
0.02 mg
0.07 mg
Linearity
±0.02 mg
±0.03 mg
±0.15 mg
±0.25 mg
Stabilization time
(Typical at FAST)
Approx. 7
seconds
Approx. 7
seconds
Approx. 7
seconds
Approx. 3
seconds
Sensitivity drift
±2 ppm/°C (10 °C to 30 °C)
Operating environment
5 °C to 40 °C, 85 %RH or less (No condensation)
Display refresh rate
5 times/second or 10 times/second
Environmental sensors
Resolution: Temperature ±1.5 °C, Humidity ±10 %, Air pressure ±10 hPa
Applicable range: 5 °C to 40 °C
Display mode
gram
Communication function
RS-232C
External calibration weight
30 g
20 g
10 g
5 g
2 g
1 g
30 g
20 g
10 g
5 g
2 g
1 g
300 g
200 g
100 g
50 g
20 g
10 g
300 g
200 g
100 g
50 g
Size of loading area for
weighing mm
Weighing pan:
φ
34 mm
Weighing pan:
φ
50 mm
External dimensions
259(W) x 466(D) x 326(H) mm
Power supply &
AC adapter type
Power consumption: Approx. 30VA (supplied to the AC adapter )
Confirm that the adapter type is correct for the local voltage and power
receptacle type.
Mass
Approx. 3.6 kg
The operating environment does not include excessive change of ambient temperature, humidity,
vibration, drafts, magnetic fields and static electricity.
The internal weight may change due to corrosion or other damage caused by the operating
environment, or due to aging. Check the internal weight periodically and correct the internal weight
value if necessary.
When placing the tare weight within the standard range, if using the smart range, the weighing
sample can be weighed within the precision range.
This is a figure measured by loading and unloading a 1g weight on the device using the automatic
machine.
The repeatability for 10 gram within the weighing capacity.