JANUARY 2010 - TD 220a DEPLOYMENT GUIDE, RDCP600
Page
87
Reliable solutions
5.6.2
Optional Feature: Updating the Surface Cell Setting
NOTE! This section holds an optional configuration item that may or may not be relevant for
your instrument. Read through the item and check if it is relevant for you.
This procedure is only required if your current Deployment Configuration uses a Surface Cell.
Earlier versions of the software did not setup the position of the surface cell properly if the cell
itself was edited afterwards. To correct the surface cell setup do as follows:
1.
Tap the
Deployment Setup
item in the main menu.
2.
Tap the
Edit
menu in at top of the overview window and select
Profile
.
3.
In the
Profile
window, tap the
Columns
page.
4.
Uncheck the
Use surface cell
in the upper right corner. The surface cell should now
disappear from the list and the check box should be unchecked.
5.
Check the
Use surface cell
. The surface cell
should now be included in the list with a red
icon. The numbers under the column
Start
and
Stop
should now be symmetrical with
Start
having a negative value, refer to Figure 5-5.
6.
Press
OK
until the Deployment Setup
application closes and the system flashes the
new setup.
Figure 5-5 Surface Cell Update