JANUARY 2010 - TD 220a DEPLOYMENT GUIDE, RDCP600
Select the Pressure page to set the update interval
of the pressure measurements.
Choose between
Read only at recording instance
,
Read continuously
, and
Use custom update interval
for the different settings. When choosing the latter
also specify the update interval.
When the deployment is configured without
surface referred columns, only the first option is
available.
When surface referred columns are configured, it
will be an advantage to update the pressure more
often than the current measurements are taken for
optimum positioning of the cells; an increased
pressure update results in increased current drain.
Figure 1-18 Timing configurations,
Pressure
1.8
Data Storage
In the
Device
page of the Data storage configuration,
select the medium and the path for where the
instrument should store the data.
The default setting is
MMC
to a directory named
RDCP
.
Make sure the
Store data to file
is checked, or else
the instrument will not store data at all.
NOTE! Hardware for Network Connections is not
available.
Figure 1-19 Data Storage, Device