JANUARY 2010 - TD 220a DEPLOYMENT GUIDE, RDCP600
Page
23
Reliable solutions
The attenuated pulse types may be used to reduce power consumption. This setting may be used
when the range is not set to full range. The optimum setting depends on e.g. the pulse length and
scattering conditions, but as a general guide with a 2m pulse and Low Power: -3dB can be used
up to 25m; -6dB can be used up to 15m; -12dB can be used up to 10m.
If the Auto Pulse setting is checked, the instrument will automatically select the proper
attenuation based on the signal strength. Auto pulse will use either 0 dB, -3 dB or –6 dB
Attenuation.
The
Columns
page is used to configure each
column.
Add a new column by tapping the
Add
button.
Select an existing column and edit or delete that
specific column as required by tapping the
Edit
or
the
Del
button.
The option
Use Surface Cell
is only available if a
quartz pressure sensor is installed.
Information about each configured column is found
at the bottom of the page.
Figure 1-10 Profile Configuration Columns