JANUARY 2010 - TD 220a DEPLOYMENT GUIDE, RDCP600
Page
83
Reliable solutions
5.5.3
Instructions for updating the Flash Registry
1.
NOTE! This step is only required if the instrument is set to start the Recorder immediately
when power is applied! Else proceed to step 2.
After the Recorder application window has
appeared, tap the
Stop
button and Close the application by tapping the Close marker in the
top right corner.
Tap the
Menu
button and select
Diagnostics
.
2.
In the Diagnostics suite scroll down and tap the
Update Registry
.
3.
When the Registry Update window appears, tab the
Update Registry
button. The instrument
will then show a small window, which tells you that the instrument is flashing the updated
Registry for approximately 2 seconds; next, the instrument will reset. It now takes the usually
10 seconds before the LCD again lights up with the main menu.
IMPORTANT!
Sometimes the updating takes place in two steps. If this is the case, the instrument will do
another flashing of the registry after it has rebooted due to the action above. It will then reset
again for a second reboot. When the instrument comes up after the second reboot, it will be
ready for operation
.
Do not switch off the power during flashing of the registry!
Note! After updating RDCP 600 Image with New Registry, make sure to update the Sensor
Board, refer to page 84.
Figure 5-1 Diagnostics Menu
Figure 5-2 Update Registry