JANUARY 2010 - TD 220a DEPLOYMENT GUIDE, RDCP600
When using our conductivity sensor 4019 you must
check the
Sensor is active
to activate the sensor.
Tap
OK
in the top right corner to save new
settings.
Hydrostatic Pressure Sensor Configuration:
In the
Calibration page
, set the maximum Pressure
Range for the Quartz Pressure sensor 3187.
Tap TEST for a performance check of the sensor.
IMPORTANT! Do not check or uncheck the option
Calibration coefficients are stored in the pressure
sensor.
If the option
calibration coefficients are stored in
the pressure sensor
is not checked, calibration
coefficients must be entered manually; tap Edit and
enter the sensor values.
If the option
calibration coefficients are stored in
the pressure sensor
is checked, the sensor
coefficients are automatic read from the sensor
Eeprom. Edit calibration coefficients if necessary
(not recommended): Select the calibration
coefficients to be edited. Tap edit and enter the
new value.
IMPORTANT! We do not recommend
that you edit the coefficients since the sensor
calibration will then not be valid.
Figure 1-32 Configuring the Conductivity
cell 3619
Figure 1-33 Sensor Configuration, Quartz
Pressure Sensor