8 Maintenance
8.1 Introduction
The device was designed, developed and tested according to the latest technical knowledge.
We have set the product life to 7 years. In addition, the availability of spare parts is
guaranteed by us within a period of 10 years. However, in order to ensure that everything
works properly, we have made it possible for you to carry out a visual check of the status
indicators from the outside.
A T T E N T I O N
There is no need for the laser user to perform routine or service work within the laser
system. All adjustments and calibration work that require the protective housing to
be opened are carried out by trained service personnel. This also includes cleaning
and cleaning the optics within the laser.
8.2 Safety Check (STK)
Once within 24 months, the laser must be subjected to a safety check (STK) by trained
personnel. The execution of the STK and any faults are to be noted in the device book.
Please note that a medical device book and a regular technical safety check are not required
in every country. Note local requirements and laws.
Scope of Safety Check (STK)
Visual Check
•
laser marking (laser class, max. power, wavelength)
•
Information signs / warning signs; properly and completely attached
•
Instructions for use / medical device book
•
Condition of the supply line
•
Condition of the goggles / protective device
•
Overall condition
Functionality Check
•
Foot switch
•
Beam guidance system / coupling / decoupling / pilot laser
•
Check operating foil (touch screen)
Check the necessary monitoring / safety display and signaling unit
•
Emission controls (acoustic, visual)
•
Laser stop switch (check for function)
•
Interlock device (check for function)
Electrical Safety Check
•
According to IEC 62353 or DIN VDE 0751, Part 1