10.3 Electromagnetic Immunity (2)
The laser is intended for use in the electromagnetic environment specified below. The customer or the
user of the laser should ensure that it is used in such an environment.
Immunity tests
IEC 60601-Testlevel
EMV standard
Electromagnetic environment
–
guidelines
Conducted
disturbance variables,
induced by high-
frequency fields
3 V/m
150 kHz to
80 MHz
IEC 61000-4-6
Portable and mobile radio devices
should not be used closer to the laser
(including the lines) than the
recommended protective distance,
which is calculated according to the
equation applicable to the
transmission frequency.
Recommended protecty distance:
d
= [ 1,17 :
V
1 ]
√
P
d
= [ 1,17 :
E
1 ]
√
P
for 80 MHz to 800 MHz
d
= [ 2,33 m/V * vP
for 80 MHz to 2,7 GHz
with
P
as the nominal power of the
transmitter in watts (W) according to
the transmitter manufacturer's
specifications and d as the
recommended protective distance in
meters (m).
Die The field strength of stationary
radio transmitters should be lower
than the compliance level at all
frequencies according to an on-site
examination.
Faults are possible in the vicinity of
devices with the following symbol.
Radio frequency
electromagnetic fields
in the immediate
vicinity of wireless
communication
devices
3 V/m
80 MHz to 2,7 GHz
IEC 61000-4-3