User Manual FOX 514
September 2021
File: User Manual_FOX 514_rev3
15
4 Safety Information
4.1 General
The FOX 514 Laser is a precise working instrument for medical applications. The system
has been carefully developed and tested through intensive testing before shipping. In order
to offer you and the operating personnel all possible protection, we recommend that you read
this section of the operating instructions carefully.
The FOX 514 is a laser of laser class 4 according to EN 60601-2-22, EN 60825-1
respectively.
Class 4 in the standard describes high-energy lasers and therefore special measures have
to be considered before use in order to ensure safe and trouble-free working with the device.
In particular, it is important to protect the eyes and skin of the operator, the patient and third
parties. Laser safety goggles must be used for eye protection.
The following explanations are not exhaustive. All users of laser devices should enclose
applicable legal regulations and provisions with the device and inform the staff accordingly.
If the device is operated outside Germany, provisions of the American National Standard
Office ANSI Z136.3-2018 "American National Standard for the Safe Use of Lasers in Health
Care Facilities" and ANSI Z136.1-2014 "American National Stand
ard for the Use of Lasers”
can be achieved.
This manual is limited to the operation, maintenance and control of the device. The manual
is not a guide for the treatment of diseases that can be treated with laser.
With regard to the devices supplied, such as fibers or probes, the safety, operating and
maintenance instructions in the corresponding manuals must be observed. A.R.C. Laser
GmbH cannot be held responsible for damage or damage resulting from improper use.
The warranty for the device expires if the laser has been opened (even partially), modified
or repaired by unqualified personnel.