User Manual FOX 514
September 2021
File: User Manual_FOX 514_rev3
43
8.4 Cleaning of Accessories
8.4.1 Laser Safety Goggles
The cleaning instructions provided with the laser safety goggles must be respected.
In general, never use chemicals for cleaning that could damage the coating fo the glasses.
Always use soft cloths that do not leace scratches on the surface
8.4.2 Applicators
In most cases, the applicators (sterile and non-sterile) are single-use products and must
not be reprocessed. Please follow the instructions in the manual for Fibers and Probes.
8.4.3 Foot switch
The general cleaning instructions (see above) must be respected in any case.
When cleaning the foot switch, prevent the penetration of water or cleaning agents. Never
place the foot switch in water or cleaning agent, this could lead to a short circuit.
8.5 Error Detection
The FOX 514 was developed and designed as a modular system. In addition, all its
components had undergone a rough shake test as well as temperature testing. In case
of an error, you can use this manual to isolate the problem and, in some cases, even
solve it yourselfIf the problem persists, please contact your local A.R.C. Laser authorized
dealer for technical support.
However, we do hope that the device will work flawlessly for many years.
When an error occurs in the system, the error will be displayed along with a symbol
to identify the error.
Common error might be:
•
Battery operation: Make sure the battery has been charged and the laser is switched
on.
•
Interlock: The interlock plug should be fully inserted into the device; the door contact
switch should be closed respectively.
•
The red laser-Stop button on the top of the device needs to be pulled up.
8.5.1 System Check
In general, a system self check runs after turning on the device, intended to check all
important functions. If a failure occurs, you will be informed immediately and the error detail
shows up on the screen.
We distinguish between two types of errors:
Dynamic errors
i.e. errors that can be easily resolved or resolved after ordering a spare
part from A.R.C. Laser GmbH.
Such errors may include:
•
Laser STOP button is pressed down
•
Interlock plug missing, disconnected or defective
•
Foot switch pressed down when turning on device, or defective
•
Fiber inserted during system start
Static errors
are errors which require contacting A.R.C. Laser GmbH service department or
the local authorized dealer. For these usually an error message will appear on the display
and can assist to identify the problem when calling for service.