Page 64
User Manual
(ESD) IEC
61000-4-2
± 2 kV, ± 4
kV, ± 8 kV,
± 15 kV /air
± 2 kV, ± 4
kV, ± 8 kV, ±
15 kV /air
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/
burst IEC
61000-4-4
± 2 kV for
power
supply lines
± 1 kV for
input/outpu
t lines
± 2 kV for
power
supply lines
± 1 kV for
input/outpu
t lines
Mains power quality should be
that of a typical professional
healthcare facility environment.
Surge IEC
61000-
4-5
± 0.5 kV, ± 1
kV line(s) to
line(s)
± 0.5 kV, ± 1
kV, ± 2 kV
line(s) to
earth
± 0.5 kV, ± 1
kV line(s) to
line(s)
± 0.5 kV, ± 1
kV, ± 2 kV
line(s) to
earth
Mains power quality should be
that of a typical professional
healthcare facility environment.
Voltage dips,
short
interruptions
and
voltage
variations on
power supply
input
lines IEC
61000-4-11
<5% UT
(>95% dip in
UT) for
0,5 cycle
40% UT
(60% dip in
UT) for 5
cycles 70 %
UT (30% dip
in UT) for
25 cycles
<5% UT
(>95% dip in
UT) for 5 s
<5% UT
(>95% dip in
UT) for
0,5 cycle
40% UT
(60% dip in
UT) for 5
cycles 70 %
UT (30% dip
in UT) for
25 cycles
<5% UT
(>95% dip in
UT) for 5 s
Mains power quality should be
that
of a typical professional
healthcare facility environment.
Power
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a
location in a typical professional
healthcare facility environment.
Radiated RF
IEC 61000-4-3
Table 9 in
IEC-60601-
1-2 2014
Table 9 in
IEC-60601-
1-2 2014
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Heron™ IOS system,
including cables, than the
Conducted RF
IEC 61000-4-6
3V 0.15-
80MHz
3V 0.15-
80MHz