![3Disc Heron IOS User Manual Download Page 42](http://html.mh-extra.com/html/3disc/heron-ios/heron-ios_user-manual_442192042.webp)
Page 42
User Manual
Based on the left and right-side bite scans, the Heron™ IOS will automatically
reconstruct the alignment of the full Maxillary and Mandibular arches in occlusal
mode.
Note
: The
Automatic Bite Alignment
may take a minute or two to process.
When
the Automatic Bite Alignment
is
completed, the 3D reconstructed image
is displayed as shown.
Manual Bite alignment
2.
Carrying Out A Manual Bite Alignment
The
Manual Bite Alignment
tool enables you to
manually
select
specific location points on the Maxillary and Mandibular arches.
This can be useful if the Auto-Alignment is unsuccessful, due to
insufficient data.
To Carry out a
Manual Bite Alignment
:
A.
Align Mandibular-Paired Points
1.
Click on the
Manual Alignment
icon in the Bite Alignment tools
menu at the bottom of the Bite Alignment window.
2.
For the HeronClinic™ to recognize the previously
scanned Mandibular and Maxillary arches,
scan
a 3-4 teeth and 3-4mm gum segment
on the
LEFT side molar/pre-molar region in Occlusion.
You will be prompted to select a mandibular-
paired point on the bite scan segment.
3.
Click
Select Point
and choose a point on the
mandibular arch of the LEFT side bite scan.