PhySys Edition SD Service Manual
Rev. B
20.08.2013
Page 47/52
Service Manual PhySysSD_RevB_20.08.2013.doc
Error /Error
code(xxxxx)
Description
Error /Error
code(xxxxx)
Description
-91
FLASH checking error
100
RAM test failed
-92
runtime RAMCODE checking error
101
a critical variable was changed unexpectedly
-93
IRQ timing error
102
an EM lookup table is corrupted
-94
Ticker tasks not executed properly
103
bootloader: there is no Application in FLASH
-95
Idle Task
104
a supervised variable's value is out of range
-96
watchdog test failed
105
105 the firmware is running on the wrong
hardware (EM/US)
-97
there was an unexpected reset by external
watchdog
-98
there was an unexpected reset by internal
watchdog
10.4.4. Cable Control Problems
When increasing the current to the patient, the message “Cable control” appears.
Remedy:
¾
Check, if the electrode cables are correctly connected
¾
Inspect the electrode cables, electrode clamps
¾
Replace electrode cables if they are contaminated or corroded
¾
Inspect the suction-wave cables