Figure 23
8.3.1.8. Replacement of the battery
¾
Opening the rear part of housing (see 8.3.1.1)
¾
Remove battery using a screw driver (see Figure 24)
¾
Insert the new battery
¾
After replacing the battery close the rear part of housing
¾
Connect the device to the power source and turn on the device
¾
In the maintenance window of the configuration menu set the current RTC (Real time clock) again and
confirm these inputs
¾
Close the configuration window and restart the device again