8.3.1.10.Replacement of power supply board
¾
Opening the rear part of housing (see 8.3.1.1)
¾
Remove the front part of housing from the lower part
¾
Disconnect all cables from the power supply board
¾
Unscrew the 6 screws (PT EJOT 3,0x8mm) and take the board out
¾
Place the new board onto the lower part of housing and screw it by using the 6x screws
¾
Connect all cables again to the board
¾
Mount the front part and rear part to the lower part of housing
¾
Run a function check (use the Safety check in chapter 12)
Figure 26