PhySys Edition SD Service Manual
Rev. B
20.08.2013
Page 48/52
Service Manual PhySysSD_RevB_20.08.2013.doc
10.5. Problems without Error message
Cable control error even if the electrodes already checked and correctly fixed.
¾
In this case it could be an electrical contact problems between the electrode cables which be fixed by screws at
the water separator and the patient connector. Please unscrew the electrode cables and ensure that the metallic
surfaces of spring and patient connector free of plastic remainders. After this screw the electrode cables again
and check the connection by applying stimulation current at the suction electrodes.
No pressure at the suction cups even if the pump running.
¾
Please check the setting therapy pressure at the Vacuum control window.
¾
Check if the suction electrodes are correctly attached.
¾
Check all silicone tubes for leakages or faulty mounting.
¾
If the problem still there, please contact Service department (Zimmer MedizinSysteme)
Pressure is too low.
¾
Please check the setting therapy pressure
¾
Check the silicone tubes and the suction cables for leakages
¾
With an external pressure sensor check the internal pressure sensor. To do this, please compare the adjusted
pressure with the measured pressure at the water separator. The max. allowed variance is ±10%.
The pulsation mode is not possible.
¾
Check the Pulsation modus in the therapy window
¾
If it possible to set the therapy pressure on a higher level but not possible to relieve the pressure, that means that
the valve is defective or the valve connection at the main board is not correctly plugged.
The pump does not stop running even the vacuum at the suction cups already reached. The pressure level on
the vacuum panel is still at minimum.
¾
Check if the silicone tubes correctly attached to the pressure sensor. Other ways the pressure sensor is probably
defective
The pump does not run even if the communication (PhySys Edition SD/ VacoP) is there (vacuum panel on the
therapy window) and the therapy pressure already set on a higher level.
¾
Please check the pump cable and pump connector on the main board.
¾
Check the cable of the water level sensor and the sensor connector on the main board.
¾
If the problem is still there, the pump is defective and must be change.
VacoP device does not run and the vacuum panel on the therapy window is not displayed even if all cables are
connected.
¾
Check all VacoP fuses (1x power inlet fuses, 1x fuse on the main board) and change the fuses if it necessary.
¾
Check the Jumper on the main board of the VacoP device (JP1001, JP1002 for the 230V power supply and
JP1003, JP1004 for the 110V power supply.
¾
If the problem is still there, please contact service department of the Zimmer MedizinSysteme.
There is no stimulating current at the suction cup electrodes!
¾
Please check all cable connections for channel 1 and 2
¾
Start a new therapy with the VacoP device.