Application information
5
Page 7
Applicator
The applicator is located at the end of the fiber-optic cable. The laser beam is
emitted from the front tip of the applicator. The outlet is protected against dirt
and damage with a lens.
Point the applicator only on the area to be treated. Never place it down
outside the calibration sensor.
Foot switch
The foot switch is used to initiate the laser beam. Press the foot switch only if
the applicator is already pointed on the part of the patient's body that is to be
treated. An acoustic signal sounds while the laser beam is being emitted.
Locate the foot switch in such a way, that it cannot be operated accidentally
or by unauthorised persons.
Emergency stop
button
The emergency stop button enables the immediate interruption of device
operation by disconnecting the power supply.
To interrupt operation, press the emergency stop button until it clicks and the
device shuts down.
To resume operation, return the button to its original position by turning the
red knob in the direction shown by the arrows.
Calibration sensor
The calibration sensor enables the amount of laser output to be measured
and adjusted.
The calibration sensor is also used for the applicator to be placed after
treatment and when not in use. This protects the applicator from dirt and
damage.
Safety glasses
All persons present in the treatment room (patient, therapist, support staff)
should wear proper protective eyewear.
Only use goggles with an optical density OD > 3 at 810nm and 980nm and a
translucence of at least 20% in the visible range. The glasses must be heat-
and UV-resistant and comply with the requirements of EN 207.
Spacer (optional)
There are two spacers of different lengths and treatment surfaces available
that will hold the laser head at a defined distance from the skin.
Protective sleeves
Protective sleeves are available when using the device without the spacers.
They can be placed on the applicator to enhance patient comfort.
Silicone spacer
(optional)
For the treatment of skin diseases, optional spacers made of silicone are
available; these must be placed on the applicator to avoid contamination.
Note:
For more additional instructions on the use, cleaning and disinfection of the
space, please refer to the spacer instruction manual.
Target beam
Opton has a target beam that provides information about the direction of the
laser beam.
The target beam is not the same diameter as the laser beam and is not an
indicator of the surface area being treated.