Functional tests
18
Page 28
Note
Ensure that all persons in the treatment room are wearing safety glasses.
Functional test
1.
Make sure that the applicator is fully inserted into the calibration sensor.
2.
Turn the device on.
3.
Enter the key code.
4.
Press the "Calibration" button.
5.
Activate the laser by pressing "Start".
6.
Perform the calibration as described in Chapter 10. The message
"Calibration OK" in the status line confirms that the laser unit is in good
condition.
7.
Press the emergency stop button. The appliance must switch off
immediately. Return the button to its original position by turning the knob in
the direction shown by the arrows.