Calibration
10
Page 18
Calibration is used to control the amount of laser beam emitted. It should be
performed each day before the first use of the laser.
Calibration
procedure
1
Close the doors of any interlock plug that has been installed.
2.
Make sure that the laser wand is in the calibrator.
4.
Activate the laser by pressing the "Start" button.
5.
The maximum laser power is automatically entered.
6.
Press the foot switch for the specified duration.
During calibration, the elapsed time is shown in the bar graph in window (2).
Calibration results
The status bar (1) will display the calibration results.
1. Calibration OK:
Laser is operating normally
2. Limited accuracy:
There is a deviation in the laser output.
This suggests contamination or age-related changes to the laser system. The
device can continue to be used, but maintenance should be performed within 4
weeks.
3. Calibration failed:
The laser output is outside the allowable range.
The device must not be used. Put the device out of operation and contact your
service representative.
A repeat of the calibration can be performed. Before calibration can be
repeated, the calibrator must first cool down. This will be indicated in the status
bar.