Operation instruction
9.1 Performing treatments
9
Page 14
Starting the program
The treatment screen will open automatically after the home screen has
displayed for four seconds or by pressing the "Start" button.
Set operating modes
Press "Parameters" (1) to open the "Parameters" window.
Here you can select options for the different operating modes.
Set the values using the arrow buttons (2).
Press the check mark (3) to accept the changed settings.
Press the X (4) to cancel the changes.
Operating modes
The following operating modes can be selected:
- no pulsing
- single pulse
- serial pulsing
Available options in
the various operating
modes
Single pulse:
Pulse time 0.5 – 5 seconds
Frequency
Duty cycle
Serial pulsing:
0.5 Hz, 0.6 Hz, 0.7 Hz, 0.8 Hz,
0.9 Hz, 1 Hz, 2 Hz, 3 Hz, 4 Hz,
5 Hz, 10 Hz, 25 Hz, 50 Hz
1:1–1:8
Setting the intensity
By turning the intensity controller, you can select the level of the desired output
(5). The setting is in increments of 0.1 W.