Manufacturer's EMC declaration
21
Page 32
Electric medical devices such asOpton are subject to special precautions with respect to electromagnetic compatibility (EMC)
and must be installed and operated according to the user manual or EMC instructions included with the device.
Portable and mobile RF communications equipment (such as mobile telephones) can affect electrical medical devices.
Opton may only be operated with the original parts specified in the list of items included in the delivery and available OEM
accessories. The operation of the device with other parts may result in increased emissions or likelihood of malfunctions!
Guidelines and manufacturer's declaration concerning electromagnetic emissions
The Opton device is intended for use in the electromagnetic environment specified below. The customer or user of
Opton should ensure that it is used in such an environment.
Emissions measurements
Conformity
Electromagnetic environment guidelines
RF emissions per CISPR 11
Group 1
The Opton device uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and it is unlikely that nearby
electronic equipment will be impaired.
RF emissions per CISPR 11
Class B
The Opton device is suitable for use in all
facilities, including residential ones, that are
directly connected to the public mains that
also supplies residential structures.
Emissions of harmonics per IEC 61000-3-
2
Class A
Emissions of voltage fluctuations / flickers
per IEC 61000-3-3
Conforms
Table 201 per EN 60601-1-2
The device must not be used adjacent to or stacked directly on other devices. If it must be used in the vicinity of or stacked
directly on other devices, the device should be monitored to verify that it is operating properly in this set-up.