Figure 10
Appendix D: Tibial Central Cone
Implant Assembly
(cont.)
Optional Tibial Augment Assembly
If an augment is being utilized, position the augment
on the underside of the tibial baseplate implant and
align the screws with the thread holes. Use the 3.5
mm hex driver to start both augment screws in a
clockwise direction. Once both screws have been
started, align the augment with the perimeter of the
tibial baseplate implant as desired. Tighten the screws
until the augment is rigidly attached to the tibial
baseplate implant (Figure 10).
Note:
The tibial central cone may need to be lifted
and rotated in order to gain access and tighten the
augments screws.
Note:
Augments must be attached after the stem
has been assembled. The 10 and 15 mm augment
thicknesses will block access to the 2 mm set screw
anteriorly.