55 |
Persona
®
Revision
Knee System
Surgical Technique
Appendix C: Femoral Central
Cone Preparation
(cont.)
Silver/Gold Broach Preparation
Attach the smallest sized broach that best fits the size
of the femoral defect to the DR handle by pulling back
the lever and locking into place (Figure 6).
Note:
The central femoral cone provisionals and
broaches are used for both “Left” and “Right”
components and must be flipped in the correct
orientation to accurately represent the 6 degree
angle of the femoral stem housing (Figure 6a).
Slide the assembly over the extension post and align
the anterior surface of the broach parallel to the
anterior cut of the femur. To broach to the appropriate
depth, the “0” line should align at or proximal to the
distal surface of the femur. If distal augments are
required, the +5 and +10 markings on the DR handle
represent the appropriate depth for broaching. If the
broach is unstable or the defect is unfilled, repeat with
sequentially larger broaches until the desired fit is
achieved (Figure 7).
To remove the DR handle/broach, rotate the lever to
the unlocked position and use the slap hammer to
remove the broach from the femur (Figure 8).
Figure 7
Figure 6a
Note:
If the broach is not proximal enough in the
femur, it may cause the femoral component to sit
proud.
Note:
The silver broaches are designed to be
broached up to 20 mm proximal of the femoral box.
Gold Broaches
The gold cone broach is a coupled device. Broaching
deeper in the proximal tibia will drive the provisional
stem deeper into the tibial canal.
If the gold broach needs to sit more proximal in
the femur, remove the stem extension post/stem
provisional construct and proceed with freehand
broaching using the DR handle and the gold femoral
cone broaches. The stem extension post/stem
provisional construct must be removed. Broaching will
drive the provisional stem deeper in the femoral canal
which could potentially cause the femur to fracture.
Note:
The extra small silver cone broach and gold
cone broach are meant for progressive broaching
only. There is not a corresponding Trabecular Metal
Cone implant offered in this size.
A
PPE
N
D
IX
C
5 mm Femoral Distal
Augmentation
No Femoral Distal
Augmentation
Distal Surface
5 mm Augment Cut
Figure 6
Figure 8