15 |
Persona
®
Revision
Knee System
Surgical Technique
Tibial Sizing and Positioning
(cont.)
Ensure optimal tibial coverage and rotation and
remove the TASP handle. Mark tibial rotation and the
keel fins. The medial 1/3
rd
of the tibial tubercle can
be used as a reference for tibial rotational alignment
(Figure 23).
Note:
It is important to mark the fins to align for the
keel. Tibial rotation cannot be adjusted after the
keel has been prepared.
Remove the reamer using the slap hammer or
T-handle, if necessary.
Figure 24
Figure 23
Tibial Stem Housing Preparation
Straight Stem Preparation Only
If the planned stem size has a diameter less than
16 mm, the bone must be cleared to accommodate
the tibial stem provisional. The 16 mm tilt reamer can
be used for this purpose. Drive the tilt reamer in the
tibial canal until the thin tibial line is level with the
most proximal tibial resection (Figure 24).
Thin Tibial Depth Line