38 |
Persona
®
Revision Knee System
Surgical Technique
Implant Assembly
(cont.)
Femoral Implant Assembly
Prior to assembly of the final construct, place the
femoral implant on a rigid table protected by a soft
surface. Use the 2 mm hex driver to loosen the set
screw to remove the taper cap from the femoral
implant. Ensure the set screw in the stem housing is
backed out in order for the stem to seat fully (Figure
98). The femoral impaction saddle may be used to
aid in assembling femoral augments and the stem to
the femoral implant. Assemble the femoral impaction
saddle with the appropriate “Left” or “Right”
designation into the underside of the impaction base.
The arrow on the femoral impaction saddle will align
with “L” or “R” on the impaction base depending on
if a left or right knee is being used (Figure 99). Place
the femoral implant on the femoral impaction saddle
(Figure 100).
Figure 99
Figure 101
Figure 98
Optional Femoral Augment Assembly
If a distal or posterior augment is being utilized,
position the augment on the femoral component
and align the screw with the threaded hole. If both
distal and posterior augments are being utilized,
assemble the distal augment onto the femoral
implant first, then the posterior augment. Use the
3.5 mm hex driver to start and align the screw in
a clockwise direction. Align the augment with the
perimeter of the femur as desired. Tighten the
screw until the augment is rigidly attached to the
femoral component (Figure 101).
Figure 100