45 |
Persona
®
Revision
Knee System
Surgical Technique
Once the cement has cured, remove any excess
cement before final bearing assembly. Confirm that all
cement has been removed from the proximal surface
of the tibial baseplate implant especially posterior near
the locking mechanism.
Two Stage Cementing Technique
(cont.)
The knee should be placed in extension for the cement
to cure. A TASP can be used to assist with compression
on the cement. Insert a TASP 2 mm thicker than
the final bearing in order to increase the pressure
on the cement (Figure 122). If the knee cannot be
straightened out at the time of cementing the femur,
constant pressure should be exerted in order to obtain
the best possible compression on the cement.
Note:
Any remaining cement will prevent proper
insertion of the TASP assembly.
Figure 122