64 |
NexGen MIS LPS-Flex Mobile Implant System
Surgical Technique
Techniques for 17 mm and 20 mm
Articular Surface Assembly
A secondary locking screw is required for the 17 mm-
and 20 mm-thick articular surface components.
Once the articular surface has been implanted, place
the secondary locking screw (packaged with the
articular surface) through the hole in the articular
surface (Figure 14e).
Attach the MIS counter torque wrench to the
articulating surface stop on the tibial plate (Figure 14f).
Tighten the MIS counter torque wrench by turning the
knob clockwise. Ensure that the hooks are flush with
the top surface of the tibial plate. Do not over-tighten.
Alternatively, the LCCK tibial plate wrench that
matches the size of the implant may be used. Place the
end of the wrench over the tibial base plate. Ensure
that the wrench is in line with the base of the tibial
base plate.
Use the LCCK deflection beam torque wrench with
the 4.5 mm hex driver bit to torque the screw to 95
in-lbs. Do not overtorque or undertorque.
Loosen the MIS counter torque wrench by turning the
knob counter clockwise.
Figure 14e
Figure 14f