62 |
NexGen MIS LPS-Flex Mobile Implant System
Surgical Technique
Femoral Component
Knee in 70°-90° flexion.
Place the collateral retractor
laterally, an Army-Navy retractor anteriorly, and a rake
retractor on the meniscal bed medially.
Place a layer of cement on the underside of the
prosthesis and in the holes drilled in the femur.
Attach the femoral impactor/extractor to the femoral
component. Insert the femoral component onto the
distal femur by translating the component laterally
until the lateral peg aligns with the drill hole in the
lateral femoral condyle. Take care to avoid scratching
the implant component surfaces. Disposable, plastic
tibial plate protectors may be temporarily inserted
onto the tibial base plate to protect the implant
surfaces during insertion of the femoral component.
Remove the tibial plate protector after the femur is
seated. Be sure that soft tissue is not trapped beneath
the implant. Use a mallet to impact the component
until fully seated.
Remove the femoral impactor/extractor, and the
retractors. Use the femoral impactor and mallet to
make sure the femoral component is fully impacted.
Check the medial and lateral sides to make sure the
femoral component is fully impacted. Remove any
excess cement in a thorough and consistent manner.