42 |
NexGen MIS LPS-Flex Mobile Implant System
Surgical Technique
Step Nine: Prepare the Tibia
Note:
If using the headless pins or small-head
holding pins, predrill using the 3.2 mm bone screw
drill.
Select the appropriate size MIS LPS-Flex mobile
broach plate (Figure 9a).
Base the selection first on achieving good mediolateral
coverage, and then anteroposterior coverage.
Verify that the femoral and tibial component sizes will
be compatible. If there is a femoral/tibial mismatch,
consider using the fixed bearing system.
Assemble the LPS-Flex mobile broach and trialing
plate. Position the trialing plate onto the broach plate
(Figure 9b) so that the peg on the under side of
the trialing plate mates with the anterior hole on
the proximal surface of the broach plate. Align the
peg and hole to prevent bending the peg. Snap the
plates together tightly (Figure 9c & 9d).
Note: If the
plates are not tightly snapped, it will interfere with
trialing.
Figure 9b
Figure 9d
Figure 9c
Figure 9a