16 |
NexGen MIS LPS-Flex Mobile Implant System
Surgical Technique
Assemble the Guide (cont.)
Insert the correct right or left tibial cut guide into
the adjustable rod and rotate the thumb wheel
counterclockwise until the threads engage.
Continue to rotate the thumb wheel until the guide
is approximately midway through its range of travel.
This will allow the depth of the tibial resection to be
adjusted after the assembly is secured to the bone
via the tubercle anchor.
Note:
The grooves on the stem of the tibial cut
guide represent 2 mm increments (Figure 1c).
Attach the correct right or left tubercle anchor onto
the corresponding side of the adjustable rod. For a left
knee, the left anchor is inserted into the right hole. For
a right knee, the right anchor is inserted into the left
hole. Be sure that the etched line on the side of the
tubercle anchor aligns with the corresponding etched
line on the anterosuperior face of the adjustable rod
(Figure 1d).
Note:
The tibial cut guide and tubercle anchor
are available in left and right configurations. If the
incorrect tubercle anchor is used, the cut guide
will not fully retract into the adjustable rod.
Figure 1c
Figure 1d
Tubercle Anchor
Содержание NexGen MIS LPS-Flex