NexGen MIS LPS-Flex Mobile Implant System
Surgical Technique
Option Two: MIS QS Notch Guide
Position the appropriate size MIS QS notch guide
onto the femur so it is flush against the resected
surfaces both distally and anteriorly. The MIS QS
notch guide will not contact the anterior chamfer.
Use the previously prepared trochlear recess and/or
the femoral post holes to position the MIS QS notch
guide mediolaterally. Use the recut guide modular peg
with MIS QS notch guides sizes C-G for mediolateral
positioning.
Secure the MIS QS notch guide to the femur with two
3.2 mm (1/8 in.) headed screws or predrill and insert
two 3.2 mm (1/8 in.) holding pins (Figure 7i). Use a
reciprocating saw to cut the sides and the base of the
intercondylar notch (Figure 7j). Then remove the MIS
QS notch guide (Figure 7k).
Figure 7i
Figure 7k
Figure 7j