Surgical Technique
Step Thirteen:
Perform a Final Trial Reduction
Insert the tibial plate provisional. Use the correct
size MIS LPS-Mobile tibial plate provisional to ensure
proper fit before implanting the final components
(Figures 13a & 13b).
Place the tibial impactor onto the provisional and
impact until completely seated (Figure 13c).
Use the femoral inserter/extractor and align the
femoral provisional onto the prepared bone, and
impact.
If preferred, the femoral provisional may be positioned
by hand.
Translate the femoral provisional laterally until the
lateral peg of the provisional aligns with the drill hole
in the lateral femoral condyle. Push the provisional
in place beginning laterally, then medially. Be sure
that soft tissue is not trapped beneath the provisional
component.
Figure 13a
Figure 13b
Figure 13c