Electromagnetic immunity for non-life-supporting ME equipment
and ME systems
The device S100 / OPMI picois intended to be operated in one of the electro-
magnetic environments specified below. The customer or the user of the de-
vice S100 / OPMI pico is responsible for ensuring that it is operated in such an
environment.
Interference re-
sistance
Tests
IEC 60601
test level
Compliance level Electromagnetic environment -
Guidelines
Portable and mobile radio communication
equipment should not be used closer to the de-
vice S100 / OPMI pico, including cables, than the
recommended safety distance that is calculated
using the equation applicable to the transmis-
sion frequency involved.
Recommended safety distance:
Conducted RF dis-
turbances as de-
scribed in
EN 61000-4:-6
3 V
actual value
150 kHz to 80 MHz
3V
Radiated RF distur-
bances as de-
scribed in
EN 61000-4:-3
3 V
actual value
80 MHz to 2.5 GHz
3 V/m
where P is the output power rating of the trans-
mitter in watts (W) according to the transmitter
manufacturer's specifications and d is the rec-
ommended safety distance in meters (m). The
field strength of stationary transmitters as deter-
mined by a site survey should be
a
less than the
compliance level in all frequency ranges.
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol.
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is influ-
enced by absorption and reflection by structures, objects and persons.
d
1 2 P
,
=
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
d
1 2 P
,
=
d
2 3 P
,
=
Содержание S100/OPMI pico
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...