OPERATION
Routine Mode
LSM 510
4-80
B 40-051 e 07/98
You can now select Z-stack limits (symmetrically
to the central Z-frame), the number of steps
and the interval size.
Further modifications of the stack parameters can
be performed as follows:
Click on the <Use New Parameter> button.
It is possible to select from fixed stack sizes of
10 µm, 20 µm, 30 µm, 50 µm and 100 µm.
After selection of the stack size, choose <Make
Z Cut> or <Next>.
Start acquisition of the stack by clicking on
<Start> in the "Routine Mode - Image
Acquisition" window.
Fig. 88
Fig. 89
Содержание LSM 510 Inverted
Страница 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Страница 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Страница 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Страница 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Страница 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Страница 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Страница 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Страница 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Страница 46: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Performance Features of the LSM 510 LSM 510 3 12 B 40 051 e 07 98...
Страница 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Страница 58: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Performance Features of the LSM 510 LSM 510 3 12 B 40 051 e 07 98...
Страница 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Страница 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...