INTRODUCTION TO LASER SCANNING MICROSCOPY
Performance Features of the LSM 510
LSM 510
3-10
B 40-051 e 07/98
(3) Transmitted-light illumination
a)
The illuminator support contains a security circuit, which activates a shutter preventing laser light from
reaching the stand when the support is moved to back. A complementary shutter built-in the stand
prevents laser light from reaching the eye pieces during scanning mode.
b)
The illuminator support is equipped with a rotatable polarizer. The Axiovert description contains the
adjustment for DIC mode during conventional observation.
For scanning transmitted light DIC mode the polarizer in the transmitted light support works like an
analyzer and must be adjusted in such a manner, that direct laser light will be blocked.
The conventional analyzer slider in the stand is not allowed be located in the beam path because of the
laser light already is polarized.
c)
On the illuminator support as an option there is mounted a LSM 5 software controlled switching
mirror fully motorized. Alternatively the light is directed to the LSM 5 T-light detector or enables
conventional transmitted-light observation.
d)
The focusing screen for conventional transmitted-light is located in a support in front of the halogen
lamp housing.
e)
Further information to halogen lamp and condensers you will find in the Axiovert operating manual.
(4) Reflected light fluorescence
All Axiovert fluorescence accessories exceptional the reflector slider can be used.
Further information you will find in the Axiovert operation manual.
(5) Imaging optics
Optovar sliders are not usable.
The analyzer for conventional DIC mode will be operated from the right side and is located just below
the nosepiece.
Use of sliders with auxiliary objects (473704/14) is not possible.
(6) Photo equipment
The stand doesn´t have an integrated SLR-port, but microscope cameras, as described in the Axiovert
operation manual, can be used.
(7) TV adaption
The TV port aside and the tubes can be used as described in the Axiovert operation manual.
The TV interface side port or base port can only be used with TV adapters 44.
Содержание LSM 510 Inverted
Страница 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Страница 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Страница 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Страница 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Страница 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Страница 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Страница 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Страница 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Страница 46: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Performance Features of the LSM 510 LSM 510 3 12 B 40 051 e 07 98...
Страница 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Страница 58: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Performance Features of the LSM 510 LSM 510 3 12 B 40 051 e 07 98...
Страница 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Страница 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...