OPERATION
Image Optimization
LSM 510
4-100
B 40-051 e 07/98
For overlaying fluorescence and transmitted-light
images, click on the <Transmission> button on the
Beam Path and Channel Assignment
panel.
The transmitted light PMT will be switched active.
Of course, all other transmitted light applications
like
phase contrast
differential interference contrast (DIC)
polarization contrast (Pol)
darkfield
can also be performed.
Image optimization must be performed as
described in the previous chapter.
For the generation of images in reflection, the main dichroic beam splitter must be a neutral-density
filter.
Standard equipment contains a neutral-density filter with a division ratio of 80 to 20 % (at 543 nm).
Fig. 4-108
Содержание LSM 510 Inverted
Страница 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Страница 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Страница 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Страница 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Страница 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Страница 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Страница 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Страница 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Страница 46: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Performance Features of the LSM 510 LSM 510 3 12 B 40 051 e 07 98...
Страница 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Страница 58: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Performance Features of the LSM 510 LSM 510 3 12 B 40 051 e 07 98...
Страница 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Страница 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...