2 Requirements | 2.4 Network Connectivity
ZEISS
Battery Charger
MAINS IN
Battery
Inverter
LOAD
Normal Mode
Backup Mode
Fig. 4: Block diagram of offline UPS
2.4 Network Connectivity
In order to provide remote diagnostics, ZEISS recommends connecting the computer to the inter-
net in a safe and secured method.
As a guideline, ask the ZEISS service representative about recommendations for data security pol-
icy on ZEISS Systems.
Cable type
CAT-6a
Connector type
RJ45 (IEC 60603-7 8P8C Modular connector)
Cable length
Site dependent – customer supplied. Not to
exceed total length of 100 m.
Protocol
Ethernet 10BASE-T/100BASE-TX/1000BASE-T
2.4.1 ZEISS Predictive Service
ZEISS Predictive Service is used for condition monitoring by systematic retrieval of relevant instru-
ment health information and its server based processing. The goal is to detect deviations in the in-
strument performance before they impair the user’s result or even result in unplanned downtime.
For details, ask the ZEISS service representative about recommendations for data security policy
on ZEISS Systems.
For predictive service, we require the above network connectivity and predictive service will oper-
ate under the parameters shown in the table below.
Ports
443
Protocols
HTTPS, secure websocket
URL of ZEISS server
predictive-service.zeiss.com
IP addresses
52.174.243.245
Connection security
TLS 1.2 with AES 256 bit
Supported proxy settings for installation
§
No authentication
§
Basic authentication
§
Digest authentication
§
NTLM authentication
12
Installation Requirements ZEISS Lightsheet 7 | en-US | Rev. 1 | 349500-8186-000