ZEISS
2 Requirements | 2.6 Gas Supplies
2.6.1 Compressed Air
Active damping tables require compressed air to isolate the system from floor vibrations. The nec-
essary compressed air can be either generated by a compressor (please ask your ZEISS sales and
service representative) or taken from a gas cylinder or from an in-house supply system.
Passive damping tables/systems typically do not require compressed air.
The compressor is connected to a separate mains socket. It runs occasionally and can introduce vi-
bration and unwanted noise to the site. Compressed air is used to operate several valves and the
auto leveling system.
Connection hose
¼ inch O.D. or 6 mm O.D.
Pressure
0.35 - 0.6 MPa (3.5 - 6 bar)
Typical flow rate
Approx. 12 l/min @ 0.6 MPa pressure during
air leveling system inflation
Quality
Oil-free
System connection
Quick exchange connector. One piece is de-
livered with the Microscope System.
The gas regulators should be supplied by the user if the air supplied is from the building facilities
or from gas bottles.
Due to acoustic noise and vibrations, where possible, the compressor should be installed in a sep-
arate room close to the workstation. Make sure there is enough room for air circulation to the
sides and at least 100 cm to the back of the compressor. Refer to the manufacturer's documenta-
tion for further details.
2.6.2 CO₂ and other Gas Supplies
For systems making use of gases such as CO
2
or O
2
in their processes or experiments, please take
note to safely handle the gases.
Make sure the gas bottles are stored safely in or outside of the building according to local safety
and fire safety regulations. The gas regulators should be supplied by the user.
To prepare for the gas connections, it is recommended to discuss with the installation engineer to
have converters available (see below) to be able to connect different size hosing depending on
what sizing units are standard in your country (6 mm to 4 mm and if possible ¼ inch to 6 mm).
Connection hose
¼ inch O.D. or 6 mm O.D.
Max. pressure
2 bar
Installation Requirements ZEISS Lightsheet 7 | en-US | Rev. 1 | 349500-8186-000
15