manualshive.com logo in svg
background image

ZEISS

2 Requirements | 2.2 Location Requirements

2.1.4 Vibration

Vibrations can be caused e.g. by heavy-duty machinery installed on the same floor or even in the
same building as well as transport facilities operated nearby. Depending on the floor stability and
construction, even walking in the room or in the hallways may affect the image quality.

Allowable vibration values (under operating conditions on system table):

§

Less than Vibration Class VC-C

, 12.5 μm/s (IEST RP 12 and ISO 10811)

Reference for floor measurements:

§

Please conform with the attenuation specifications of the proposed system table to ensure the
VC-C specification is met on the table.

§

For the NEWPORT Vision IsoStation series of damping tables recommended by ZEISS:
Less than 63 µm/s for 0 to 4 Hz and less than 125 µm/s for 4 Hz to 1000 Hz

§

These recommendations are based on an attenuation of -10 dB starting from 4 Hz and over
-26 dB above 10 Hz. Low frequency vibrations (swaying) is recommended to be less than half
of the 4 Hz limits.

Requirements for measuring vibrations:

§

Three positions should be measured.

§

The positions need to cover a triangle over the area where the microscope will be installed.

§

At each position, measurements in X, Y and Z need to be taken.

§

X is along the front of the microscope.

2.2 Location Requirements

The Microscope System must only be operated in closed rooms. It is recommended to install the
Microscope System in a dimmed room where artificial illumination, sunlight or other light sources
cannot interfere with image acquisition. The Microscope System should not be installed near radi-
ators or windows with direct sunlight. Compliance with the installation requirements of the Mi-
croscope System and the availability of the requested supplies is the responsibility of the customer
and has to be readily available at the time of installation.

Installation site

Exclusively inside buildings

Operating range

Max. 2000 m above sea level

§

Do not place the Microscope System near windows with direct sunlight or radiator heating
systems which could change the temperature of the system.

§

Do not position power supplies or racks under the system table because the heat dissipated
can cause instability of the Microscope System through the expansion of the steel table.

§

Harmful gases and fumes that may be used on the Microscope System must be safely ex-
tracted from the Microscope System and site in accordance with local safety regulations.

§

The Microscope System must not be set up in areas with potential danger by explosives.

2.2.1 Space Requirements

Recommended room size

Min. 4.0 m x 5.0 m x 2.3 m

Service and safety clearance area

Min. 0.5 m around the Microscope System

Customer arranged computer workspace

Min. 0.8 m x 1.2 m x 0.76 m

Entrance

Min. 0.8 m wide

Installation Requirements ZEISS Lightsheet 7 | en-US | Rev. 1 | 349500-8186-000

7

Содержание Lightsheet 7

Страница 1: ...Installation Requirements ZEISS Lightsheet 7 ...

Страница 2: ...es for archiving purposes shall remain unaffected thereby Any viola tions may be prosecuted as copyright infringements The use of general descriptive names registered names trademarks etc in this document does not imply that such names are exempt from the relevant intellectual property laws and regulations and therefore free for general use This shall also apply if this is not specifically referre...

Страница 3: ...gory 10 2 3 2 UPS Requirements 11 2 4 Network Connectivity 12 2 4 1 ZEISS Predictive Service 12 2 5 Safety and Radiation Information 13 2 5 1 Laser Safety 13 2 5 2 Biosafety Level of the Laboratory 14 2 6 Gas Supplies 14 2 6 1 Compressed Air 15 2 6 2 CO and other Gas Supplies 15 2 7 Cooling System 16 2 8 Transport and Storage 16 2 8 1 Weight and Sizes of the Transported Goods 16 2 8 2 Weight and S...

Страница 4: ...emperature Air quality The customer is responsible for ensuring that the conditions specified in this document are met and that the required equipment is available before delivery and installation A ZEISS service repre sentative will gladly assist you in examining the installation conditions and suitable measures 1 1 Further Applicable Documents Please take also note of the following documents Dow...

Страница 5: ...0 dB Z class 200 to 500 Hz less than 55 dB Z class 2 1 2 Air Conditioning and Quality The following nominal power usage for the Microscope System and possible accessories is shown in Watts To maintain a more constant room temperature the size of the air conditioning should be well matched to the size of the room and the heat generated The stable air conditioning of the room is critical to achievin...

Страница 6: ...llution occurs Occasionally temporary conductivity caused by condensation is to be expected Example A circuit used in an of fice environment The circuitry inside a computer would fall in this cate gory 3 Conductive pollution occurs or dry nonconductive pollution occurs that becomes conductive due to condensation that is to be expected Example Circuitry that is exposed to outside air but will not b...

Страница 7: ...tem must only be operated in closed rooms It is recommended to install the Microscope System in a dimmed room where artificial illumination sunlight or other light sources cannot interfere with image acquisition The Microscope System should not be installed near radi ators or windows with direct sunlight Compliance with the installation requirements of the Mi croscope System and the availability o...

Страница 8: ...asurements in mm For possible changes in pref erential layout please discuss with your ZEISS representative The placing of the user table can vary depending on available space and cable lengths Data cable max 30 m 500 ca 1600 750 800 500 900 1200 approx 2300 1200 5 4 9 8 6 7 2 1 3 850 incl corner pieces 1000 incl corner pieces 500 Fig 1 Space requirement for Lightsheet 7 with a storage and data an...

Страница 9: ...E 220 240 V AC 10 100 125 V AC 10 Nominal frequency 50 60 Hz Main Power Plug Local mains plug will be supplied Power consumption 220 240 V 800 VA 100 125 V 750 VA Storage PC Max 400 VA Max current 3 5 A at 220 V 8 A at 100 V Additional building PE The system must be connected to a building earth point at all times Additional mains power 3 x Standard local 230 or 120 V mains sock ets Required for t...

Страница 10: ...ation of motors generators or the switching of reactive circuit components Random transients on the other hand are often caused by lightning and electro static discharge Ensure that the possibility of transient voltages on the mains line are minimal to prevent damage to the system Max transient voltage in V Nominal Voltage VAC Category I Category II Category III 50 330 500 800 100 500 800 1500 150...

Страница 11: ...ements If the site requires a UPS Uninterruptable Power Supply please ensure the KVA rating of the UPS meets the ratings of the Microscope System and all third party items that may also be connected e g water chiller air compressor sample prep equipment EDS or WDS etc This will ensure that the complete system will continue to function during a blackout or brownout long enough to allow the user to ...

Страница 12: ...d for condition monitoring by systematic retrieval of relevant instru ment health information and its server based processing The goal is to detect deviations in the in strument performance before they impair the user s result or even result in unplanned downtime For details ask the ZEISS service representative about recommendations for data security policy on ZEISS Systems For predictive service ...

Страница 13: ...safety regula tions The following types of lasers are currently intended for use Using other lasers than indicated be low requires prior coordination with ZEISS The performance values of the directly integrated mul tiphoton lasers correspond to the rated laser power Laser Wavelength nm Class Typical power ex fiber mW solid state laser SSL 405 3B 20 solid state laser SSL 405 3B 50 solid state laser...

Страница 14: ...er some do exist in a small number of places in the US and around the world As the highest level of biological safety a BSL 4 lab consists of work with highly dangerous and exotic mi crobes Infections caused by these types of microbes are frequently fatal since there are no available treatments or vaccines Two exam ples of such microbes include Ebola and Marburg viruses Tab 3 Biosafety Level 2 6 G...

Страница 15: ...user if the air supplied is from the building facilities or from gas bottles Due to acoustic noise and vibrations where possible the compressor should be installed in a sep arate room close to the workstation Make sure there is enough room for air circulation to the sides and at least 100 cm to the back of the compressor Refer to the manufacturer s documenta tion for further details 2 6 2 CO and o...

Страница 16: ...he weight information on the package and on the shipping document Where possible the original packaging must be used for shipping or transport Fork lift and a hand pallet truck For on site transport and unloading a fork lift and a hand pallet truck are necessary Ensure all hallways and corners are wide enough to be passed with the hand pallet truck Please check the location requirements for door a...

Страница 17: ...ves some indication on the approximate weight and sizes of the unpacked items Component Length mm Width mm Height mm Weight kg System table small incl corner pieces 900 1000 750 850 810 870 130 Main system module Light sheet 7 800 450 500 75 LB Rack Lightsheet 700 600 550 80 Cooling unit INNOVATEK 3156 01 500 130 190 7 5 Info The tables for the computers are provided by the customer ZEISS recommen...

Страница 18: ... international guidelines and standards the 21 CFR 1040 10 Performance Standards for light emitting products laser products applies for the USA The following EMC user notice is for Korea only 기종별 사용자안내문 A급기기 업무용방송통신기자재 이기기는업무용 A급 전자파적합기기로서 판매자또는사용자는이점을주의하시기바라 며 가정외의지역에서사용하는것을목적으 로합니다 European and International Directives Standards For more information on ISO and CSA certifi cates CE Declarations o...

Страница 19: ...h requirements are yet to be completed ZEISS can be requested to assist with the checking of the environmental conditions Environmental conditions in and around a system can change with time due to various changes to the building new equipment being installed or other reasons Should the environmental condi tions change due to changes to the building or other potential external influence after a si...

Страница 20: ...Radiation Information X Ray Laser Safety Biosafety Level of the Laboratory Cooling Water Supply Cooling liquid and or chemicals Exhaust Line Gas Supplies Compressed Air Nitrogen Additional Gas Transport and Storage Place date Name Signature The above signature conveys that the Installation Requirements have been read and understood The customer further understands that some of these requirements a...

Страница 21: ......

Страница 22: ...Zeiss Promenade 10 07745 Jena Germany phone 49 1803 33 63 34 fax 49 3641 64 3439 info microscopy de zeiss com www zeiss com microscopy Installation Requirements ZEISS Lightsheet 7 en US Rev 1 Modifications reserved Printed in Germany ...

Отзывы: