SETUP REQUIREMENTS
Lattice Lightsheet 7
Power Requirements
ZEISS
04/2021 V_01
000000-2324-960
57
The Lattice Lightsheet 7 system must be connected to the AC network via the Power Supply Unit (PSU) by a
country-specific mains power cable (Fig. 36).
The 6 point power strip must be separately connected to the AC network by further power cables.
Power connectors and fuses of the PSU, the laser module and the Lattice Lightsheet 7 are displayed in Fig. 36,
and Fig. 37, respectively.
Line voltage
230 V AC (±10 %)
120 V AC (±10 %)
Line frequency
50 / 60 Hz
50 / 60 Hz
Lattice Lightsheet 7 system
incl. lasers
Max. current
1 phase at 4.5 A
1 phase at 9 A
Power
Power consumption
900 VA (continuous operation, maximum)
280 VA (standby operation)
0.025 VA (off mode)
900 VA (continuous operation; maximum)
260 VA (standby operation)
0.011 VA (off mode)
Power plug
Country specific
Country specific
Class of protection
I
I
Ingress protection mode
IP 20
IP 20
Overvoltage category
II
II
Pollution degree
2
2
Heat dissipation
700 W max.
700 W max.