SAFETY INFORMATION
Lattice Lightsheet 7
User Interface - Mounting and Dismounting Components
ZEISS
04/2021 V_01
000000-2324-960
41
Heated Tubing Package
The Heated Tubing Package (Fig. 24) connects the gas output of the Gas Mixer to the Humidifying Column
and to the Heated Chamber on the microscope stage. The gas tubing carrying humidified gas is heated up
to avoid condensation. For this purpose, a heating wire is inserted in the tubing.
1
Electrical cable for the heating of the Humidifying Column
6
Heated wire (split off from electrical cable)
2
Gas tubing to the Heated Chamber
7
Insulating jacket
3
Merging point for heating wire into the tubing
8
Mixed gas output white
4
Gas output from Humidifying Column
9
Mixed gas output yellow
5
Gas input to Humidifying Column
Fig. 24
Heated Tubing package
Arrows indicate flow direction of the gas mixture.
Electrical connection
1.
Cable to connect the Electrical plug socket of the Humidifying Column (Fig. 23/
3
) to the Gas Mixer Heater
socket (Fig. 22/
13
).
2.
Heat wire electrical split off from the electrical cable into both tubing from Heating Column tubing
connector red (output) (Fig. 23/
1
) and tubing to heated chamber.