SAFETY INFORMATION
Lattice Lightsheet 7
User Interface - Mounting and Dismounting Components
ZEISS
04/2021 V_01
000000-2324-960
27
8.
Release the Luer-Lock couplings (Fig. 11/
9
and
/
10
).
9.
Remove the old meniscus lens.
10.
Place the new meniscus lens in the meniscus lens
holder.
11.
Connect the Luer-Lock couplings.
12.
Insert the meniscus lens into the incubator
13.
Turn the meniscus lens until the three-point
support is in place.
14.
The orientation groove (Fig. 12/
7.1
) must point
to the left.
15.
Reinsert the sample mounting frame.
16.
Reinsert the incubator lid.
17.
Lower the transmission arm.
5
Meniscus lens holder
7
Meniscus lens
9
Luer-Lock coupling
10
Luer-Lock coupling
Fig. 11
Luer-Lock couplings
7.1
Orientation groove
Fig. 12
Orientation groove