59
000000Ć1150Ć839 IOLMaster 11.02.2004
2. FineĆshifting the measuring cursor
Note:
This manipulation always requires previous zooming in!
Proceed as described above under Item 1, but use the right mouse
button for dragging the measuring cursor. This way the automatic
peak detection is deactivated and the white point may be positioned
at any place over the measuring curve.
When you release the mouse button, the current axial length and the
new SNR will be calculated and displayed.
This kind of manipulation is advisable with closely adjacent double or
triple peaks.
Also in this case the recalculated axial length is denoted in the display
field by an asterisk (*).
Explanation:
Even if you undo the manipulations with the measuring cursor (by
moving it back to the automatically found maximal peak) and the
measured value agrees with the original one, the asterisk after the
measured value will remain. This is to indicate that the curve has been
manipulated!
The described manipulations of the measuring cursor may be performed
both in axial length measurement mode (after the current individual
measurement) and in post measurement editing mode.
Evaluation of ALM results