000000Ć1150Ć839 IOLMaster 11.02.2004
PostĆrun editing of axial length measurements
The results of axial length measurements must be interpreted on the
basis of the signalĆtoĆnoise ratio and the appearance of the graphs
(
cf.Ă
"Example signal curves of axial length measurements"
, p.53
).
The illustrations below do not contain the video image.
Fig. 43 Presentation of the graph of the fifth axial length measurement (23.00 mm;
SNR: 6.6) on the right eye (shown without video image)
SNR categories
While the system is internally calculating the axial length from the
interference signal, it automatically analyzes the SNR.
SNR > 2.0 --> The measured value is valid.
SNR in the range 1.6 ... 2.0 --> The measured value is not reliable
(Fig. 44)
The signalĆtoĆnoise ratio may be low for the following reasons:
Ć Dense medial opacity along the visual axis,
Ć restless patients,
Ć alignment of instrument to patient eye is not optimal,
Ć very high ametropia (> 6 D),
Ć corneal scars,
Ć retinal pathology.
Fig. 44 Non reliable value
Evaluation of ALM results